Skip to content
  • The objective of this test is to verify whether the DEEPSOUND equipment can successfully detect underlying defects within the composite material.

small-image Specimen Front

  • Front Side

small-image Specimen Back

  • Back Side

  • Measurements are conducted by placing the probe directly onto the sample without a wedge. The goal is to determine if points of varying sizes (Point #1 to Point #3) can be distinctly identified in the PAUT S-scan.

small-image 2.25 MHz PAUT Probe

  • Equipment: 2.25 MHz PAUT Probe

small-image Marking Signals

  • Marking Each Signal Point

  • Actual Thickness of Point #1: Approximately 7.00 mm
  • Detected Position of the Reflected Signal: 6.13 mm
  • Observation: The S-scan clearly displays the ultrasonic signal reflecting off the intermediate rubber layer.

small-image Point #1 Result

Point #1 S-Scan

  • S-Scan Image of Point #1

  • Actual Thickness of Point #2: Approximately 11.00 mm
  • Detected Position of the Reflected Signal: 10.38 mm
  • Observation: A clear reflected signal is visible at the designated depth.

Point #2 Result

Point #2 S-Scan

  • S-Scan Image of Point #2

  • Actual Thickness of Point #3: Approximately 23.00 mm
  • Detected Position of the Reflected Signal: 22.31 mm
  • Observation: The S-scan clearly displays the ultrasonic signal reflecting off the intermediate rubber layer at its maximum thickness.

small-image Point #3 Result

Point #3 S-Scan- S-Scan Image of Point #3


  • The test successfully confirmed that explicit reflected signals could be detected at each designated inspection point within the composite specimen.
  • The inspection was conducted using a probe with a frequency of 2.25 MHz. For optimal results in highly attenuating materials, we recommend utilizing a probe with a frequency lower than 2.25 MHz.
  • Because composite specimens inherently cause severe attenuation of ultrasonic waves, images can appear somewhat blurred. Measuring alongside a known defect-free reference sample is strongly advised for comparative analysis.
  • When inspecting the maximum thickness at Point #3, identifying the signal reflecting off the far back-wall proved challenging. (Recommendation: Increase the transmission voltage of the probe to proactively compensate for severe signal attenuation.)
  • Ultimately, this ultrasonic testing comprehensively verified the presence of internal structural changes. Employing a lower frequency probe is expected to yield noticeably sharper and cleaner images.